| Generator | ||
| kW Output | 65 kW or 80 kW | |
| High Frequency generator with AEC | Output: 65 kW according to EN 60336 at 100 kV Ripple rate < 1 kV at 100 kV 100 kV @ 650 mA 400 mA @ 150 kV |
|
| Radiographic parameters | From 40 to 150 kV, by step of 1 kV (or displacement in the range by slider), accuracy ± (3% + 1 kV) From 10 to 650 mA, accuracy ± (4% + 1 mA), 19 values (10, 12.5, 16, 20, 25, 32, 40, 50, 65, 80, 100, 125, 160, 200, 250, 320, 400, 500, 650) From 1 ms to 10 s, accuracy ± (2% + 0.1 ms), 38 values 1, 2, 3, 4, 5, 6, 8, 10, 12, 16, 20, 25, 32, 40, 50, 65, 80, 100, 125, 160, 200, 250, 320, 400, 500, 650, 800 ms and 1, 1.25, 1.6, 2, 2.5, 3.2, 4, 5, 6.5, 8, 10 s From 0.1 mAs to 500 mAs |
|
| Fluoroscopy parameters | From 40 to 120 kV, by step of 1 kV, accuracy ± (3% + 1 kV) From 0.5 to 5 mA (low dose) and up to 7 mA (higher dose) |
|
| Operating mode | 3 points (kV, mA, s) 2 points (kV, mAs) 1 point (kV, AEC) |
|
| AEC | 3 chambers | |
| Remote Control Table | ||
| SID | Motorized and continuous from 110 to 180 cm, 6 cm/s | |
| Column angulations | Motorized, +/-40°, 10°/s | |
| Tube rotation | Manual, +/-180°, with electromagnetic brake and mechanical stops every 90° that permits, with table in vertical position, to makes exposures on stretcher or on wall stand (dependant on the configuration) | |
| Tomography | ||
| Layer thickness | 1.5 mm (at 40°), 3 mm (at 20°), 6 mm (at 10°) | |
| Maximum speeds per angle | 0.5 s (10°); 1 s (20°); 2 s (40°) | |
| Patient Support | ||
| Table height | 64 to 93 cm, 6 cm/s | |
| Table top | Flat, 225x81 cm, 60 cm radio transparent | |
| Absorption | 0,4 mm eq Al | |
| Patient load | Dynamic weight is 500 pounds or less in any direction - Static weight is 660 pounds (between 501 and 660 pounds) | |
| Lateral movements | Motorized, +/- 18 cm at 6 cm /s | |
| Longitudinal movements | Motorized, +/- 50 cm (or 70/30 cm or 120/50 cm) at 10 cm /s | |
| Tilting | -25°/+90°, 6°/s | |
| FPD longitudinal travel | 135 cm, 14 cm/s | |
| Patient coverage | Head to Toe, > 2.5 m | |
| Grid | ||
| Focus | 110 cm and 180 cm | |
| Ratio | 10:1, 40 l/cm | |
| Table Console | ||
| Soft keys | Tube 0 position, table 0 position, longitudinal tabletop movements, lateral tabletop movements, auto-collimation, auto-positioning and lighting | |
| Joysticks | SID, column angulations, tilting, variable height, carriage, tabletop lateral movements and collimator shutters | |
| IR remote control | All the table movements are available | |
| Touch screen | ||
| 3 table position quick access | Stretcher, parking and chest | |
| 3 fluoroscopy field zooms | 14", 12", 9" | |
| Display | Table height, angulations, SID, tilting angle, collimator size | |
| Accessories | ||
| Footrest | ||
| Adjustable stool | ||
| 2 handles | ||
| 2 winches | ||
| 1 belt | ||
| X-ray Tube | ||
| Presetting of x-ray tube features | Loading ratings, cooling ratings, starting voltages, starting times, maximum current limitation | |
| Safety and protection disposals for x-ray tube | The electronic calculation of load by software indicates limits to the tube with message and forbids x-ray emission Housing temperature control and display of available Heat Units in % Overheated IGBTs detection Housing oil pressure control in series with x-ray tube heat safety |
|
| Max exposure voltage | 150 kV | |
| Focal spot | 0.6 and 1.2 mm | |
| Output | 40 and 100 kW | |
| Anode angle | 12.5° | |
| Anode rotation | 9000 rpm | |
| Time of anode rotation | 1.6 s | |
| Anode dimension | 102 mm | |
| Inherent filtration | 0.7 mm | |
| Anode heat storage capacity | 600 kHU (285kJ) Standard 800 kHU and 1000 kHU Optional |
|
| Anode heat dissipation rate | 1000 W (81,1 kHU /min; 60 kJ /min) | |
| House heat storage capacity | 2025 kHU (1000 kJ) | |
| Housing heat dissipation rate | 1000 W (81 kHU /min; 60kJ /min) with fan | |
| Cooling | Air | |
| High Voltage cable length | 12 meters | |
| Total filtration | IEC 60-601-3 compliant | |
| Weight | 29 kg | |
| Collimator | ||
| Inherent filtration | 1 mm Al | |
| Copper pre-filter | 0 mm - 0.1 mm - 0.2 mm - 0.3 mm motorized and automatic (pre-programmed in each protocol) | |
| Rotation | +/-30° | |
| Full field light localizer | 100 W halogen lamp, laser, timer | |
| Square collimation | Motorized, manual or automatic (pre-programmed in each protocol) | |
| DAP System | ||
| Active area | 147 mm × 147 mm | |
| Stabilization time | Approximately 6 minutes | |
| Measuring range | from 10 to 99 999 999 mGy.cm² | |
| Digital resolution | 1 mGy.cm² | |
| Sensitivity uniformity | < 6 % | |
| Accuracy | +/- 25 % | |
| Filtration Eq Al (70 kV) | 0.2 mm | |
| Transparency | > 70 % | |
| Video Camera | ||
| Video camera | Free radiation positioning | |
| Image display | On the screen of the table console | |
| Auto-Positioning | ||
Complete system operation |
Control of generator, table and imaging system from the acquisition console | |
| Examination procedure | The table position, generator parameters and collimation settings, including additional filtration and image processing parameters, are automatically set via protocols | |
| Protocols | Customizable | |
| Table Auto-Positioning | ||
| Tilting | ||
| Column angulations | ||
| SID | ||
| Lateral tabletop position | ||
| Longitudinal tabletop position | ||
| Collimation | ||
| Filters | ||
| Optional Compression | ||
| Compression strength | 0 to 130 N | |
| Available angulations | +/-10° | |

|